Care and use manual – Waters ACQUITY UPLC SEC Columns and Standards User Manual

Page 5

background image

5

[ CARE AND USE MANUAL ]

ACQUITY UPLC BEH SEC Columns

0.0

1.0

2.0

3.0

4.0

5.0

6.0

7.0

8.0

9.0

10.0 min

0.00

0.04

0.08

0.12

2

1

3

4

5

6

Figure 5. Protein Mixture on an ACQUITY UPLC BEH450 SEC, 2.5 µm column.

Table 4. BEH450 SEC Test Mix

Analyte

pI

MW

1. Thyroglobulin Dimer

4.6

1.4 million

2. Thyroglobulin, Approx 3 mg/mL

4.6

669,000

3. IgG, 2 mg/mL

6.7

150,000

4. BSA, 5 mg/mL

4.6

66,400

5. Myoglobin, 2 mg/mL

6.8, 7.2

17,000

6. Uracil, 0.1 mg/mL

N/A

112

Instrument:

ACQUITY UPLC System with Tunable UV (TUV) detector

Column:

ACQUITY UPLC BEH450 SEC, 4.6 x150 mm, 2.5 µm

Sample:

BEH450 SEC Protein Standard Mix

(P/N

186006842

)

BEH450 SEC Protein Standard Preparation:

Dissolve the BEH450 SEC Protein Standard in

500 μL of 100 mM sodium phosphate buffer pH

6.8. Once solubilized, do not freeze sample,

store at 2–8 °C for no more than a week.

Mobile Phase:

100 mM sodium phosphate, pH 6.8

Weak Needle Wash: 100% Milli-Q water

Strong Needle Wash: 100% Milli-Q water

Seal Wash:

90/10 water/methanol

Injection Type:

Full loop

Injection Volume: 2 μL

Flow Rate:

0.3 mL/min

Column Temp.:

30 °C

Detection:

UV @ 280 nm

IV. COLUMN SPECIFICATIONS AND USE

To ensure the continued high performance of ACQUITY UPLC BEH SEC,
1.7 µm columns, follow these guidelines:

a. SEC Eluent and Needle Wash Preparation

Use HPLC-grade buffers, water, and organic solvents when possible.

Filter solutions through a compatible 0.2 µm or smaller pore size
filter. The use of a sterile filtration apparatus is recommended
for buffers capable of supporting microbial growth.

Solutions that are susceptible to microbial growth should be
replaced at regular intervals to avoid column contamination.
Do NOT refill partially full SEC Eluent bottles with new Eluent.
Rather, when required use new bottle containing freshly prepared
SEC Eluent.

Select solvent inlet filters that are compatible with solutions
used, and clean or replace filters regularly when using
solutions that are susceptible to microbial growth.

b. Sample Preparation

1. Ensure that samples are free of particulates before injecting onto

the SEC column. If samples appear cloudy or turbid, they should
not be injected, as this could lead to column pressure increases.
Sample preparation such as filtration or centrifugation may be
used, if appropriate.

2. If the sample is not dissolved in the mobile phase, ensure that

the sample, solvent and mobile phases are miscible in order to
avoid sample and/or buffer precipitation.

c. Column Specifications

Shipping Solvent: 20% methanol in water

Recommended Operating Parameters

Recommended Flow Rate Range for maximum column lifetime:

0.1–0.4 mL/min

Recommended Maximum Operating Column Backpressure:

BEH125—10,000 psi for both 150 and 300 mm

BEH200—7,000 psi for both 150 and 300 mm

BEH450—4,500 psi for 150 and 7,500 psi for 300 mm

Mass Load: < 100 μg for a 4.6 x 150 mm