beautypg.com

Care and use manual, Iii. getting started – Waters ACQUITY UPLC SEC Columns and Standards User Manual

Page 2

background image

2

[ CARE AND USE MANUAL ]

ACQUITY UPLC BEH SEC Columns

II. CONFIGURING AN ACQUITY UPLC SYSTEM FOR USE IN
SEC PROTEIN SEPARATIONS

a. Calibrators

In order to obtain the best performance from your SEC column, it is
important that your ACQUITY UPLC system is properly configured. It
is recommended that only pre-cut tubing is used, and that the ID of
all connecting tubing is 0.005” or less for optimal chromatographic
performance.

Size-exclusion chromatography may require modifications to an
existing ACQUITY UPLC system. Please refer to “Size-Exclusion and
Ion-Exchange Chromatography of Proteins using the ACQUITY UPLC
System” (P/N

715002147A

) for specific recommendations that can

be obtained at www.waters.com/chemcu

The sample loop used may affect the performance of your separation.
Optimally, select the smallest volume sample loop that is required
for the application. Sample loops larger than 20 µL with the ACQUITY
UPLC BEH SEC column are not recommended.

III. GETTING STARTED

A Performance Test Chromatogram and a Certification of Analysis
(COA) are available for each shipped ACQUITY UPLC BEH SEC column.
The COA is specific to each batch of packing material and includes the
batch number and chromatographic separation obtained using defined
protein standards. This document is available upon request at
www.waters.com/chemcu

ACQUITY UPLC BEH450 SEC,
2.5 µm: 100K–1500K Daltons

ACQUITY UPLC BEH200 SEC,
1.7 µm: 10K–450K Daltons

ACQUITY UPLC BEH125 SEC,
1.7 µm: 1K–80K Daltons

Protein standards; T: 30°C; Mobile phase: 100mM sodium phosphate, pH=6.8

MW

100

1000

1,000,000

100,000

10,000

Normalized Retention Volume (Vr/VC)

0.4

1.0

0.8

0.6

Uracil (112 Da)

Allantoin (158 Da)

Angiotensin frag 1-7 (899 Da)

Aprotinin (6.5 kDa)

RNase A (13.7 kDa)

Myoglobin (17 kDa)

Ovalbumin (44 kDa)

Conalbumin (75 kDa)

Amylglucosidase (97 kDa)

IgG (150 kDa)

Thyroglobulin (669 kDa)

Blue Dextran

IgM (900 kDa)

Figure 1. Calibration Curves on ACQUITY UPLC BEH125, BEH200, and
BEH450 SEC Columns

Figure 2. Separation of Protein and Peptide Standards on ACQUITY UPLC
BEH125, BEH200, and BEH450 SEC Columns

a. eCord Installation

The eCord button should be attached to the side of the ACQUITY UPLC
column heater module. The eCord button is magnetized and does not
require specific orientation.

0.00

0.04

0.08

0.12

0.00

0.05

0.10

0.15

2

3

4

5

6

7

8 min

1

2

3

4

5

6

7

AU

AU

AU

0.00

0.02

0.04

0.06

0.08

0.10

1

1

2

2

3

3

4

5

5

6

7

7

Thyroglobulin dimer
(Approx. 1.4 million Daltons)

1.

Thyroglobulin (670K)

2.

BSA (66K)

3.

Ovalbumin (44K)

4.

Carbonic Anhydrase (29K)

5.

Myoglobin (17K)

6.

Angiotensin Frag 1–7 (899)

7.

Uracil (112)

ACQUITY UPLC
BEH450 SEC
Note: Stds 4 and 6 above
were not included in the
BEH450 Column test mix

ACQUITY UPLC
BEH200 SEC

ACQUITY UPLC
BEH125 SEC

Columns

ACQUITY UPLC BEH125, 1.7 µm

ACQUITY UPLC BEH200 1.7 µm

ACQUITY UPLC BEH450, 2.5 µm

Column
Configuration

4.6 x 150 mm

Mobile Phase

100 mM Sodium Phosphate Buffer, pH 6.8

Weak
Needle Wash

100% Milli-Q

®

Water

Strong
Needle Wash

100% Milli-Q Water

Seal wash

90/10 water/methanol

Samples:
Diluted in
mobile phase

Thyroglobulin 0.3 mg/mL

BSA 0.3 mg/mL

Ovalbumin 0.3 mg/mL

Carbonic Anhydrase 0.3 mg/mL

Myoglobin 0.3 mg/mL

Angiotensin Frag. 1–7 0.1 mg/mL

Uracil 0.1 mg/mL

Thyroglobulin 3 mg/mL

BSA 5 mg/mL

Ovalbumin 3 mg/mL

Myoglobin 2 mg/mL

Uracil 0.1 mg/mL

Injection Vol.

2 µL, Full Loop

Flow Rate

0.3 mL/min

Column Temp.

Ambient

Detection
Wavelength

UV @ 220 nm

UV @ 280 nm

Compounds: