c. Sample Load and Flow Rates
Sample load and recommended flow rates vary depending on
application. For more information, please consult an appropriate
Waters applications note at www.waters.com.
d. Removal
Gradually stop the flow and wait until decompression is completed.
This may take 1 - 2 minutes.
NEVER remove the trap column when system pressure is greater
than 200 psi since this can damage the packed column bed.
Remove the trap column by disconnecting the outlet prior to
trap inlet.
e. Recommended Storage
1. Short Term (e.g., 24 - 72 hours before reuse):
T he nanoACQUITY UPLC System should be set to flow at
300 nL/min and a composition of 50/50 water/acetonitrile.
2. Long Term (e.g., greater than 72 hours before reuse):
Store in 100% acetonitrile and NO TFA or FA.