beautypg.com

Care and use manual – Waters nanoACQUITY UPLC 2G Trap Columns User Manual

Page 3

background image

Waters Corporation
34 Maple Street
Milford, MA 01757 U.S.A.
T: 1 508 478 2000
F: 1 508 872 1990
www.waters.com

[ CARE AND USE MANUAL ]

Waters, The Science of What’s Possible, nanoACQUITY UPLC,
UPLC, and Symmetry are registered trademarks of Waters
Corporation. Optima is a trademark of Fisher Scientific. All other
trademarks are the property of their respective owners

©2013 Waters Corporation. Produced in the U.S.A.
November 2013 720004296EN VW-PDF

V. ORDERING INFORMATION

Description

Part Number

nanoACQUITY UPLC 10K-2G-V/M Trap Symmetry C

18

,

180 µm x 20 mm, 5 µm

186006527*

*This part number is a replacement for part number 186003514

c. Sample Load and Flow Rates

Sample load and recommended flow rates vary depending on
application. For more information, please consult an appropriate
Waters applications note at www.waters.com.

d. Removal

Gradually stop the flow and wait until decompression is completed.
This may take 1 - 2 minutes.

NEVER remove the trap column when system pressure is greater
than 200 psi since this can damage the packed column bed.

Remove the trap column by disconnecting the outlet prior to
trap inlet.

e. Recommended Storage

1. Short Term (e.g., 24 - 72 hours before reuse):

T he nanoACQUITY UPLC System should be set to flow at
300 nL/min and a composition of 50/50 water/acetonitrile.

2. Long Term (e.g., greater than 72 hours before reuse):

Store in 100% acetonitrile and NO TFA or FA.