[ CARE AND USE MANUAL ]

Enzymate BEH Pepsin Column

2

Protein in H O, pH 7.0, 20 °C

Protein Labeling

Labeling Quenched

Enzymate BEH Pepsin Column

UPLC Separation

Add excess D

2

O

Reduce pH to 2.5

ESI MS Analysis

Global analysis

Local analysis

Maintained at

10–20 ˚C

HDX Manager

II. HOW WAT ERS HDX T ECHNOLOGY WORKS

Waters offers a complete HDX MS system for protein higher-order
structure analysis, such as protein conformational dynamics,
protein-drug binding, protein-protein interactions including
epitope mapping, and protein aggregation. This integrated system
includes automated sample preparation with a LEAP system,
on-line pepsin digestion using the Enzymate BEH Pepsin Column,
UPLC, and QTof MS instrumentations for peptide separation
and characterization with automated data processing and
interpretation by DynamX.

ACQUITY UPLC M-Class with HDX Technology workflow:
1. Deuterium oxide (D

2

O) is added to the protein sample which is

solubilized in a physiological buffer (e.g., 100 mM phosphate
buffer, pH 7.0).

2. The exchange reaction proceeds until a defined end point or

multiple time points for rate determinations.

3. The exchange reaction is stopped by reducing the pH to 2.5

and decreasing the temperature to 0 °C.

4. Sample is injected into the injection port of the HDX Manager

which is maintained at 0 °C.

5. Proteins are digested into peptides on the Enzymate BEH

Pepsin column maintained at 10–20 °C.

Figure 1. How Hydrogen Deuterium Exchange (HDX) works in the ACQUITY UPLC M-Class System with HDX Technology.

6. The peptides are then trapped on a VanGuard™ Pre-Column.

7. The trapped peptides are then eluted from the trap column,

and separated on an ACQUITY UPLC Column at 0 °C.

III. GET TING START ED

a. Column Connection

The Enzymate BEH Pepsin Column uses eXtended Performance
[XP] or ACQUITY UPLC Column-style hardware. Therefore, the
internal geometry of the Enzymate BEH Pepsin Column end nuts
is the same as eXtended Performance [XP] or ACQUITY UPLC
Columns. This geometry is very similar to “Parker-style” depths
which is the LC industry standard. It is important, however, to
reset the fitting every time a new Enzymate BEH Pepsin Column is
installed to ensure no voids or leaks occur.

Use 1/16” PEEK tubing (0.005” internal diameter) for the
Enzymate BEH Pepsin Column inlet and outlet tubing, hand cut
to the appropriate lengths (e.g., part number WAT022995).
Alternatively, 1/16” steel tubing of similar dimensions and
lengths may also be used.