beautypg.com

Care and use manual, Iii. column specifications and use, A. sec eluent and needle wash preparation – Waters XBridge Protein BEH SEC Columns and Standards User Manual

Page 4: B. sample preparation, C. column specifications

background image

4

[ CARE AND USE MANUAL ]

XBridge Protein BEH SEC Columns and Standards

AU

0.000

0.010

0.020

0.030

Minutes

0.00

1.00

2.00

3.00

4.00

5.00

6.00

7.00

8.00

9.00

10.00

1.) Thyroglobulin dimer, 2.) Thyroglobulin, 3.) IgG FU, 4.) BSA, 5.) Myoglobin, 6.) Uracil

2

1

3

4

6

5

Figure 3: Protein Mixture separation on XBridge Protein BEH SEC 450,
3.5 µm 7.8 x 150 mm

Analyte

pl

MW

1. Thyroglobulin, 0.1 mg/mL

4.6

669,000

2. Thyroglobulin, Approx 3 mg/mL

4.6

669,000

3. IgG, 2 mg/mL

6.7

150,000

4. BSA, 5 mg/mL

4.6

66,400

5. Myoglobin, 2 mg/mL

6.8, 7.2

17,000

6. Uracil, 0.1 mg/mL

N/A

112

Table 3: BEH450 SEC Test Mix

Conditions:
Instrument:

ACQUITY TUV with Tunable UV detector

Column:

XBridge Protein BEH SEC, 450Å, 3.5 μm,

7.8 x 150 mm

Sample:

BEH450 SEC Protein Standard Mix

(

P/N 186006842

)

Mobile phase:

100 mM sodium phosphate, pH 6.8

Weak needle wash:

100% Milli-Q water

Strong needle wash: 100% Milli-Q water

Seal wash:

90/10 water/methanol

Injection type:

Full loop

Injection volume:

2 µL

Flow rate:

0.86 mL/min

Column temp.:

Ambient

Detection:

UV @ 280 nM

III. COLUMN SPECIFICATIONS AND USE

To ensure the continued high performance of XBridge Protein BEH
SEC, 3.5 µm Columns, follow these guidelines:

a. SEC Eluent and Needle Wash Preparation

Use HPLC-grade buffers, water, and organic solvents
when possible.

Filter solutions through a compatible 0.2 µm or smaller pore size
filter. The use of a sterile filtration apparatus is recommended
for buffers capable of supporting microbial growth.

Solutions that are susceptible to microbial growth should be
replaced at regular intervals to avoid column contamination.
Do NOT refill partially full SEC eluent bottles with new eluent.
Rather, when required use new bottle containing freshly
prepared SEC eluent.

Select solvent inlet filters that are compatible with solutions
used, and clean or replace filters regularly when using
solutions that are susceptible to microbial growth.

b. Sample Preparation

Ensure that samples are free of particulates before injecting
onto the SEC column. If samples appear cloudy or turbid,
they should not be injected, as this could lead to column
pressure increases. Sample preparation such as filtration or
centrifugation may be used, if appropriate.

If the sample is not dissolved in the mobile phase, ensure that
the sample, solvent and mobile phases are miscible in order to
avoid sample and/or buffer precipitation.

c. Column Specifications

Shipping solvent: 20% methanol in water

Recommended maximum flow rate and backpressure:

XBridge Protein BEH SEC 200, 7.8 x 150 mm:
4 mL/min/2,600 psi

XBridge Protein BEH SEC 200, 7.8 x 300 mm:
2.7 mL/min/3,200 psi

XBridge Protein BEH SEC 450, 7.8 x 150mm:
4mL/min / 2600psi

XBridge Protein BEH SEC 450, 7.8 x 300mm:
2.7 mL/min / 3200psi