Care and use manual – Waters XBridge Protein BEH SEC Columns and Standards User Manual
Page 3
3
[ CARE AND USE MANUAL ]
XBridge Protein BEH SEC Columns and Standards
d. Useful Functional Tests for Benchmarking LC System and
XBridge Protein BEH SEC Column
Waters recommends performing a benchmarking test upon
receipt of your column and throughout the lifetime usage. By
using a separation of common proteins with an appropriate
method, you can:
■
■
Verify the performance of the column upon receipt.
■
■
Monitor the condition of the columns for extended use.
■
■
Troubleshoot separation difficulties that may arise.
Waters BEH200 SEC Protein Standard Mix (
P/N 186006518
)
and BEH450 SEC Protein Standard Mix (
P/N 186006842
) were
specifically designed for this purpose with carefully chosen
proteins and/or peptides to provide a good representation of the
intended application.
The information below details how to successfully prepare and
use the included BEH SEC Protein Standard for benchmarking or
troubleshooting purposes. Figures 2 and 3 provide separation
conditions and results you should expect to obtain.
Buffer Preparation
Chemicals:
■
■
Sodium phosphate monobasic, monohydrate
■
■
Sodium phosphate dibasic, anhydrous
■
■
HPLC-grade water
Preparation of 100 mM sodium phosphate buffer (500 mL):
1. Weigh out 500 ± 0.02 g of water into a 500 mL beaker.
2. Weigh out 3.55 ± 0.02 g of sodium phosphate dibasic and add
to the 500 mL beaker.
3. Weigh out 3.45 ± 0.02 g of sodium phosphate monobasic and
add to the 500 mL beaker.
4. Stir for a minimum of 30 minutes then filter the solution
through a 0.2 µm filter.
5. Take the pH and record the value for reference purposes
(approximate pH: 6.8).
AU
0.000
0.012
0.024
0.036
Minutes
0.00
1.00
2.00
3.00
4.00
5.00
6.00
7.00
8.00
9.00
10.00
1.) Thyroglobulin, 2.) IgG FU, 3.) BSA, 4.) Myoglobin, 5.) Uracil
1
2
3
4
5
Figure 2: Protein Mixture separation on XBridge Protein BEH SEC 200,
3.5 µm 7.8 x 150
Analyte
pl
MW
1. Thyroglobulin, 3 mg/mL
4.6
669,000
2. IgG, 2 mg/mL
6.7
150,000
3. BSA, 5 mg/mL
4.6
66,400
4. Myoglobin, 2 mg/mL
6.8, 7.2
17,000
5. Uracil, 0.1 mg/mL
N/A
112
Table 2: BEH200 SEC Test Mix
Conditions:
Instrument:
ACQUITY TUV with Tunable UV detector
Column:
XBridge Protein BEH SEC, 200Å, 3.5 μm,
7.8 x 150 mm
Sample:
BEH200 SEC Protein Standard Mix
(
P/N 186006518
)
Mobile phase:
100 mM sodium phosphate, pH 6.8
Weak needle wash:
100% Milli-Q
®
water
Strong needle wash: 100% Milli-Q water
Seal wash:
90/10 water/methanol
Injection type:
Full loop
Injection volume:
2 µL
Flow rate:
0.86 mL/min
Column temp.:
Ambient
Detection:
UV @ 280 nM