Specifications – NTi Audio XL2 Sound Level Meter User Manual
Page 208

208
Specifications
Stabilization
Time
< 10 seconds
Integration
Time
Minimum: 1 second
•
Maximum: 100 hours minus 1 second
•
Display
Measure-
ment
Ranges
Three level ranges depending on the microphone
sensitivity with manual setting.
M2230 @ sensitivity = 42 mV/Pa
•
LOW, lower level range: 0 - 100 dBSPL
»
MID, mid-level range: 20 - 120 dBSPL
»
HIGH, upper level range: 40 - 140 dBSPL
»
M2211 @ sensitivity = 20 mV/Pa
•
LOW, lower level range: 10 - 110 dBSPL
»
MID, mid-level range: 30 - 130 dBSPL
»
HIGH, upper level range: 50 - 150 dBSPL
»
M2215 @ sensitivity = 8 mV/Pa
•
LOW, lower level range: 20 - 120 dBSPL
»
MID, mid-level range: 40 - 140 dBSPL
»
HIGH, upper level range: 60 - 160 dBSPL
»
M4260 @ sensitivity = 26 mV/Pa
•
LOW, lower level range: 10 - 110 dBSPL
»
MID, mid-level range: 30 - 130 dBSPL
»
HIGH, upper level range: 50 - 150 dBSPL
»
Residual
noise in [dB]
@ S =
42 mV/Pa
of XL2
without
measure-
ment
microphone
Frequency weighting A
•
Level range
Leq
Lpeak
LOW
4
17
MID
18
31
HIGH
43
55
Frequency weighting C
•
Level range
Leq
Lpeak
LOW
3
16
MID
17
30
HIGH
41
55
Frequency weighting Z
•
Level range
Leq
Lpeak
LOW
7
20
MID
21
34
HIGH
46
58