Capture + tolerances – NTi Audio XL2 Sound Level Meter User Manual
Page 154
![background image](/manuals/569567/154/background.png)
154
Capture + Tolerances
#Unit
dBr (relative), fixed setting
#Mode
Describes tolerance type; the XL2 allows set-
ting of the tolerance type independently of the
tolerance file.
HighLow
Utilizes upper and lower toler-
ance based on ideal reference.
High
Utilizes only upper tolerance
based on ideal reference.
Low
Utilizes only lower tolerance
based on ideal reference.
#Columns
Column headings for further tolerance data
separated in different columns:
Frequency - Min - Ideal - Max
#ATolerances Defines the tolerance band for the
A-weighted wideband level (not mandatory).
#CTolerances Defines the tolerance band for the
C-weighted wideband level (not mandatory).
#ZTolerances Defines the tolerance band for the
Z-weighted wideband level (not mandatory).
#Band
Tolerances
Defines the tolerance band for the individual
•
frequencies.
The ideal parameter can be set as “undef”.
•
The same frequency value on two succes-
•
sive rows defines a tolerance step, e.g.:
Frequency
Min
Ideal
Max
100
70
75
80
500
70
75
80
500
75
80
90
1000
75
80
90
#LevelOffset
All levels in the tolerance file are relative levels
dBr. Thus the level offset provides the relation
between the relative level and actual sound
pressure level. (not mandatory)
The XL2 allows setting of the level offset inde-
pendently of the tolerance file.