1 heated vaporizer accessory contents, 3 installation, 1 tools and supplies required – INFICON 3000 Micro GC Gas Analyzer Operating Manual User Manual
Page 95

5 - 21
PN
07
4-
51
9-
P1
C
3000 Micro GC Operating Manual
5.1.6.2.1 Heated Vaporizer Accessory Contents
Figure 5-25 Heated vaporizer accessory kit
5.1.6.3 Installation
The heated sample introduction accessory can be installed on any 3000 Micro GC
that has a heated injector.
5.1.6.3.1 Tools and Supplies Required
A power source rated for 100 to 240 V (ac), 50/60 Hz, 2 A
5/16 in. open-end wrench
Table 5-8 Heated vaporizer accessory contents
(PN G2819A-X/G2846A-X/G2858A-X)
Description
Quantity
Heated Vaporizer assembly, with
copper coil extension installed
1 ea
Copper high pressure vent tubing
1 ea
Tygon low pressure vent tubing
1 ea
Power supply and cable
1 ea
Filter disconnect assembly (includes inline filter and
disconnect fitting)
1 ea
Transfer line kit (two each transfer lines, insulation,
1/16 in. nut with front and back ferrules, and 1/4 in.
female nuts with front and back ferrules)
1 ea
Copper coil
PN G2819A-X
Heated Vaporizer
Power cable
(varies by country)
Power supply
Tygon low pressure
vent tube
Filter disconnect
assembly
Copper high pressure vent tubing
Insulation (2)
Transfer line (2)
with front and
back ferrules
(installed at factory)
14 in. female nut
PN G2846A-X