INFICON 3000 Micro GC Gas Analyzer Operating Manual User Manual

Page 136

background image

7 - 8

PN

07

4-

51

9-

P1

C

3000 Micro GC Operating Manual

4

Select the desired parameters. See

Table 7-2

for a list of parameters.

Table 7-2 Valco (VICI) sampling valve configuration parameters list

5

Click Set Config to set the configuration of the valve. See

Figure 7-12

.

Name

Description

Model

Identifies the name of the valve model. Other configuration items will be
unavailable until the Model is selected.

Number of
Pos’ns

Corresponds to the number of ports for the valve from the drop-down list.

Port

Corresponds to the communications Port for the valve from the drop-down
list.

Baud Rate Corresponds to the Baud Rate, or symbols per second, for communicating

with the valve. Consult the valve documentation.