3000 Micro GC Operating Manual

Table 5-11

shows the remote start/cancel action matrix.

5.2.1 EZ IQ Operation to Activate Remote Start/Cancel Function

For instructions on how to use an external trigger within EZ IQ, see

section 7.14,

EZ IQ Operation to Activate Remote Start/Cancel Function, on page 7-49

.

Table 5-11 Remote start/cancel action matrix

Purpose

Pin Connection

Action

Remote Start

Pin 1 and pin 2

Connect together to activate circuit

Pin 3 and pin 4

No action when open
Trigger REMOTE_START when closed

Remote Cancel

Pin 9 and pin 10

Connect together to activate circuit

Pin 11 and pin 12

No action when open
Trigger REMOTE_CANCEL when closed

NOTE: When External Trigger is defined in EZ IQ, a 5 ms contact closure in the

REMOTE_START circuit starts the 3000 Micro GC run and EZ IQ begins
data acquisition.

NOTE: When 3000 Micro GC is ready to begin a run but a REMOTE_START

signal has not been received, the READY_OUT contacts (i.e., pins 8 and
15) are closed. They remain closed until the run starts or the run is
cancelled/aborted.

NOTE: If the REMOTE_CANCEL contacts are closed, 3000 Micro GC will open

the READY_OUT relay (i.e., pins 8 and 15) to signal the connected device
that the run is terminated.