8 - 22

PN

07

4-

51

9-

P1

C

3000 Micro GC Operating Manual

Figure 8-28 Commission a standard GC Module - restart instrument countdown

10

Set the 3000 Micro GC front panel power switch to the off position.

8.2.8 Enable the Instrument in EZ IQ

Update EZ IQ with updated 3000 Micro GC hardware configuration after:

installing the replacement Standard or Performance Enhanced GC module

updating the GC firmware

For instruction on how to update 3000 Micro GC instrument configuration in EZ IQ,
refer to

section 7.2 on page 7-1

.