3000 Micro GC Operating Manual
TCD and columns turn off
(continued)
High carrier gas pressure
Ensure the carrier gas is supplying at least
78 PSI, and no more that 82 PSI.
CAUTION
Exceeding 82 PSI for the
carrier gas pressure may
result in damage to
3000 Micro GC.
Malfunctioning column or
injector
Check the flow coming out of the analytical
and reference vents on the back of the
instrument. If there is no flow on either/both
vents, contact
Malfunctioning TCD
Contact
Malfunctioning EPC valve
Contact
Inlet/column temperature not
stabilizing
Instrument requires power
cycle
Restart 3000 Micro GC as instructed in
section 6.3.2, Restart, on page 6-8
Hardware problem
Contact
Unable to download
method/error message
Instrument requires power
cycle
Restart 3000 Micro GC as instructed in
section 6.3.2, Restart, on page 6-8
Incorrect configuration
See
section 7.2, Connecting EZ IQ with
.
EZ IQ requires restart
Close EZ IQ and re-open. Restart the
computer if necessary.
Method is corrupt
Create a new method and reprogram
parameters.
Drivers are corrupt
Configure the instrument through EZ IQ
Config.
D-Board error
Contact
TCD detector 'fault'
Instrument requires power
cycle
Restart 3000 Micro GC as instructed in
section 6.3.2, Restart, on page 6-8
EZ IQ requires restart
Close EZ IQ and re-open. Restart the
computer if necessary.
Drivers are corrupt
Configure the instrument through EZ IQ
Config.
Table 9-2 Common software problems (continued)
Symptom
Cause
Remedy