beautypg.com

INFICON 3000 Micro GC Gas Analyzer Operating Manual User Manual

Page 232

background image

9 - 4

PN

07

4-

51

9-

P1

C

3000 Micro GC Operating Manual

Inverted peaks;
large, broad humps in
chromatogram

Incorrect carrier gas

Check the carrier gases and ensure that each
carrier gas is connected to the correct
3000 Micro GC inlet. See the 3000 Micro GC
webpage and verify proper gas setup. Refer to

section 6.4, Change Carrier Gas
Configuration, on page 6-9

.

Impure carrier gas

Ensure the carrier gas purity is between
99.995% and 99.9995%.
Add hydrocarbon/moisture/oxygen traps or
particulate filters to the carrier gas line to
improve purity.

Carryover peaks

If broad and sporadic peaks appear, it could
indicate carryover from a late-eluting
compound. Extend Run Time or increase
Column Temperature.

NOTE: An increase in column temperature

will change component retention
times.

Air peak on blank injection

Column head pressure low

Verify there is sufficient carrier gas pressure.
Insufficient carrier gas pressure may prevent
the inject valve from sealing fully, creating an
air leak.

Internal leak

Contact

INFICON

.

Significant negative
baseline, but peaks still
appear

Broken reference column

Contact

INFICON

.

Poor separation

Contamination

Run a bakeout to clean the columns. See

section 9.2.2.1, Column and Detector
Bakeout, on page 9-18

.

Column Temperature and
Column Head Pressure not
optimized

Reduce the Column Temperature and/or
Column Head Pressure to achieve better
separation.

Malfunctioning TCD

Contact

INFICON

.

Malfunctioning column

Check the flow coming out of the
3000 Micro GC analytical and reference
vents. If there is no flow from either or both
vents, contact

INFICON.

Table 9-1 Common chromatographic problems (continued)

Symptom

Possible Cause

Remedy