INFICON 3000 Micro GC Gas Analyzer Operating Manual User Manual
Page 232

9 - 4
PN
07
4-
51
9-
P1
C
3000 Micro GC Operating Manual
Inverted peaks;
large, broad humps in
chromatogram
Incorrect carrier gas
Check the carrier gases and ensure that each
carrier gas is connected to the correct
3000 Micro GC inlet. See the 3000 Micro GC
webpage and verify proper gas setup. Refer to
section 6.4, Change Carrier Gas
Configuration, on page 6-9
Impure carrier gas
Ensure the carrier gas purity is between
99.995% and 99.9995%.
Add hydrocarbon/moisture/oxygen traps or
particulate filters to the carrier gas line to
improve purity.
Carryover peaks
If broad and sporadic peaks appear, it could
indicate carryover from a late-eluting
compound. Extend Run Time or increase
Column Temperature.
NOTE: An increase in column temperature
will change component retention
times.
Air peak on blank injection
Column head pressure low
Verify there is sufficient carrier gas pressure.
Insufficient carrier gas pressure may prevent
the inject valve from sealing fully, creating an
air leak.
Internal leak
Contact
Significant negative
baseline, but peaks still
appear
Broken reference column
Contact
Poor separation
Contamination
Run a bakeout to clean the columns. See
section 9.2.2.1, Column and Detector
Bakeout, on page 9-18
Column Temperature and
Column Head Pressure not
optimized
Reduce the Column Temperature and/or
Column Head Pressure to achieve better
separation.
Malfunctioning TCD
Contact
Malfunctioning column
Check the flow coming out of the
3000 Micro GC analytical and reference
vents. If there is no flow from either or both
vents, cont
Table 9-1 Common chromatographic problems (continued)
Symptom
Possible Cause
Remedy