8.1.2 Shutdown/Restart

CAUTION

Avoid shutting down 3000 Micro GC using the power
switch on the front panel. This may cause a system
integrity issue.

CAUTION

If a 3000 Micro GC is powered off when the column is at a
high temperature, column contamination and
degradation may result.

Proceed as follows to shutdown or restart a 3000 Micro GC.

1

Create a shutdown method:

Turn off the Detector Filaments

Lower the Column Temperature to 45°C

Select On for Column Pressure and set to at least 20 PSI

A shutdown method is illustrated in

Figure 8-2

.

Figure 8-2 Example shutdown method

2

Download the method from EZ IQ onto the 3000 Micro GC.