beautypg.com

INFICON 3000 Micro GC Gas Analyzer Operating Manual User Manual

Page 207

background image

8 - 25

PN

07

4-

51

9-

P1

C

3000 Micro GC Operating Manual

8.3.3 Running a Low Pressure, Liquid/Particulate Containing Gas Sample

For gas samples containing liquids or particulates, use the Genie Filter Assembly
(PN G2817A). See

Figure 8-30

.

CAUTION

The Genie Filter Assembly sample conditioner is
designed to remove small amounts of liquids or
particulates. It is not suitable for samples containing
large amounts of liquids or particulates.

1

If the Genie Filter Assembly is not installed, refer to

section 5.1.3, Installing the

Genie Filter Assembly, on page 5-3

.

Figure 8-30 Genie filter sample conditioner (PN G2817A)

2

Turn 3000 Micro GC on.

3

Check that the carrier gas supply is sufficient to run all of the samples.

4

If using a sample vessel, prepare the sample.

5

Connect the sample conditioner or gas source to 3000 Micro GC.

CAUTION

For sample source pressures above 172 kPa (25 PSI),
install a pressure regulator between the sample source
and Genie filter.

6

Open the sample vessel valve and adjust the pressure to between
70 to 172 kPa (10 to 25 PSI).

7

Use EZ IQ software to load the desired method and perform the analysis.

8

When the analysis is complete, close the sample vessel valve and disconnect
the vessel from the instrument.

9

Repeat steps 4 to 8 to run the next sample.

To GC

Sample input
tubing

Vent tubing

Mounting bracket not shown