8.3.6 Running a High Pressure Gas Sample Containing C6+ Components

This procedure is used for samples from heated transfer lines, or from high
pressure sample vessels using an INFICON Heated Regulator accessory.

NOTE: The accessories for all 3000 Micro GC types are the same except for their

mounting brackets.

8.3.6.1 Using a Sample Vessel

1

If the Heated Regulator accessory is not installed, install it as instructed in

section 5.1.6, Installing Heated Sample Introduction Accessories, on page
5-17

. See

Figure 8-33

.

Figure 8-33 Sample vessel connected to heated regulator via quick disconnect fitting

(2-Channel instrument shown)

2

Turn 3000 Micro GC on

3

Turn on the Heated Regulator accessory.

4

Allow approximately thirty minutes for the Heated Regulator accessory to
stabilize at operating temperature.

5

Check that the carrier gas supply is sufficient to run all of the samples.

Table 8-3 List of heated regulator IPNs

3000 Micro GC Type

Accessory

1,2-Channel 3000 Micro GC

G2818A-X

3,4-Channel 3000 Micro GC

G2845A-X

Portable 3000 Micro GC

G2857A-X

Sample vessel

Sample vessel

stopcock

Relief valve

High pressure vent tubing

Low pressure vent tubing

Quick disconnect fitting

Stainless steel tubing (user provided)