3000 Micro GC Operating Manual

8

Select the new carrier gas type from the drop-down list. See

Figure 6-13

.

Figure 6-13 Selecting a new carrier gas type

9

Click Submit. This turns the column heaters off (to cool the columns and avoid
thermal shock) and displays the Check Status view. See

Figure 6-14

.

Figure 6-14 Check status view

10

Click Check Status to display a window which shows the current column
temperature(s). When the column temperatures are below 60

º

C, Close the

Check Status window. See

Figure 6-15

.

NOTE: Changes may be canceled up to this point by closing the browser or by

using the browser back feature.

Figure 6-15 Example check status window

Periodically refresh display

Actual column tempertures