beautypg.com

2 status, 1 check gc module parameters – INFICON 3000 Micro GC Gas Analyzer Operating Manual User Manual

Page 114

background image

6 - 6

PN

07

4-

51

9-

P1

C

3000 Micro GC Operating Manual

6.2 Status

6.2.1 Check GC Module Parameters

When EZ IQ software is used to download a method, the information is sent to
3000 Micro GC and is displayed in the Status tab. Temperatures and pressures for
all GC modules will be listed. See

Figure 6-7

.

NOTE: Temperature and pressure settings cannot be edited in this view.

Settings can only be edited within EZ IQ software.

Figure 6-7 GC module status

To check GC module parameters:

1

Open the browser and type in the 3000 Micro GC IP address.

2

Click the Status tab.

3

Compare the actual values against their set points. Do the actual values meet
their setpoints?

3a

YES—If all parameters meet set point values, proceed with

section 6.2.2,

Check the Duty Cycle, on page 6-7

.

3b

NO—If the parameter values do not meet their set point, or if there are flags
(= 00), reset the 3000 Micro GC by closing the browser and then turning
3000 Micro GC off and on. Recheck the Status tab. If the actual values still do
not meet their set points, contact

INFICON

.