3000 Micro GC Operating Manual
Proceed as follows:
1
Connect the Carrier Gas Cylinder Recharge Kit (PN PNU-2058) to the supply 
tank via the CGA-580 fitting. See 
. To avoid leaks, tighten the
connections securely.
2
Make sure the needle valve of the Carrier Gas Cylinder Recharge Kit is fully 
closed by turning the needle valve clockwise until it is firmly seated. See 
Figure 8-44 Cylinder recharging kit setup
3
Partially open the valve on the supply tank. No gas should flow.
4
Loosely connect the 1/8 in. tube from the Carrier Gas Cylinder Recharge Kit to 
the CARRIER FILL (Swagelok) port on the Portable 3000 Micro GC back 
panel. See 
.
Figure 8-45 Back of Portable 3000 Micro GC
5
Partially open the needle valve on the Carrier Gas Cylinder Recharge Kit and 
listen for gas leaking through the loose 1/8 in. fitting on the back panel. This 
purges the Carrier Gas Cylinder Recharge Kit transfer lines so that no air enters 
Portable 3000 Micro GC.
6
After the transfer lines have been sufficiently purged (about fifteen seconds), 
tighten the 1/8 in. fitting on the back panel of Portable 3000 Micro GC.