Page 116: 5 remove or replace a gc module, 6 change gc module address

6 - 8
PN
07
4-
51
9-
P1
C
3000 Micro GC Operating Manual
6.3.2 Restart
To restart the 3000 Micro GC:
1
On the Utilities tab, click Restart.
2
A countdown screen will count down three minutes.
3
After three minutes have passed, set the front panel power switch to the off
position.
4
Set 3000 Micro GC front panel power switch back to the on position.
5
Wait at least three minutes for the 3000 Micro GC to fully reboot.
6.3.3 Shutdown
On the Utilities tab, click Shutdown to shut down 3000 Micro GC. A countdown
screen will count down three minutes. After three minutes have passed, set the
front panel power switch to the off position.
CAUTION
Do not turn off the power without waiting at least three
minutes. 3000 Micro GC must update configuration files
before power is turned off. If power is turned off too soon,
the files could be corrupted, making 3000 Micro GC
inoperable.
6.3.4 Support
On the Utilities tab, click Support to go to the Product Support page.
6.3.5 Remove or Replace a GC Module
A standard GC module change can be conducted through the embedded web page
in the
Utilities tab. See
section 8.2, Remove and Replace a 3000 Micro GC
for a detailed procedure.
6.3.6 Change GC Module Address
To change the address of a GC module:
1
On the embedded web page, click the Utilities tab.
2
Click Change Module Configuration
.
3
Click Change and type the new GC module address.