HYDAC HMG 3000 User Manual
Page 37

HMG 3000
Page 37
Edition 11/2007 - V02 R01-S02
Stock no.: 669712
The sampling rate which can be set is also
dependent on the active measurement
channels:
• 0.1 ms
max. 2 analog input
channels
• 0.2 ms
max. 4 analog input
channels
• 0.5 ms
all 10 input channels
• 1.0 ms
for SMART sensors
Recording time
A drop-down list appears when this item is
selected. Using the arrow keys (
↑) and (↓)
you can select a recording time of between 1
second and 24 hours, followed by pressing
OK
.
When changing the setting, information is
shown at the bottom of the display pertaining
to the recorded data showing you the result
of your setting.
Recording option
The options available here are End when
time has elapsed
and Record
continuously until terminated
.
End when time has elapsed
means that
recording is discontinued when the time
entered in Recording time has elapsed.
Record continuously until terminated
causes the oldest measured values to be
deleted when new measured values are
added upon the time entered in Recording
time
elapsing. The time entered in
Recording time
remains as the "past" in
relation to the current time in the measured
value memory until the Stop command is
given (Stop + OK or actuate the Esc key).