HYDAC HMG 3000 User Manual
Page 35

HMG 3000
Page 35
Edition 11/2007 - V02 R01-S02
Stock no.: 669712
6.2 Recording Menu
Here the HMG enables you to perform a
measurement for a specific period of time
and save it. This recording can be rendered
as a graph or a table. To do this, proceed as
follows: To access the recording menu, go to
the Current Measured Values menu, then
select Recording in the function bar at the
bottom. In the Recording menu that follows,
select Start recording, View recording or
Delete recordings
, and confirm via OK.
6.2.1 Starting a Recording
When selecting Start recording, you
proceed to the menu in which the individual
measurement parameters which can be set
for making a recording are listed. When
selecting the first item, Recording, a drop-
down list appears in which you can choose
between Simple measurement curve,
Measurement curve with trigger
or Event
log
. Select a recording option followed by
OK
.
When selecting Simple measurement
curve
, you can make additional parameter
settings:
Measurement channels
The active measurement channels are
shown here. Confirm via OK. You now
proceed to the
Change Settings of
Measurement Channels
menu. All of the
settings you can make here are described
in detail in the Changing the Settings of
the Measurement Channels
section of
Settings Menu
. Before starting your
recording you can now decide whether
you want to switch off individual channels
(X = "off").
Important:
Don't forget to reactivate the
measurement channels after completing
your measurement, assuming they are
still needed.