beautypg.com

Dakota Ultrasonics DFX-8 plus MANUAL1 User Manual

Page 98

background image

Dakota Ultrasonics

94

3) Press the ENTER key to display the list of material types.

4) Press the

UP and DOWN arrow keys multiple times on the left button

wheel to scroll through the material list until the appropriate material is

highlighted.

5) Press the ENTER key to overwrite the material type and display the menu

items with the new material type selected.

6) Finally, press the MEAS key to return to the measurement screen and begin

taking readings.

11.6 Material Velocity

The DFX-8 has a built in feature to manually adjust the material velocity to a known
value in the calibration section of the DFX-8. It should be noted that this does not
calculate the zero value like the two point calibration option does. Therefore, if the
material velocity is changed you should check the calibration. The procedures for
adjusting the velocity are outlined below:

For quick access and editing, refer to navigating the Hot Menus page 22.

Adjusting Velocity

This manual is related to the following products: