Dakota Ultrasonics DFX-8 plus MANUAL1 User Manual
Page 94
![background image](/manuals/557264/94/background.png)
Dakota Ultrasonics
90
1) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right and the ESC key multiple times to tab left until
the PRB menu is highlighted and displaying the submenu items.
Alternatively, press the
LEFT and RIGHT arrow keys multiple times
using the left button wheel.
2) Use the
UP and DOWN arrow keys multiple times on the left button
wheel to scroll through the sub menu items until MAX PRF is highlighted.
3) Use the
LEFT and RIGHT
arrow keys on the right button wheel to scroll
the PRF options.
4) Once the desired PRF is displayed, press the MEAS key to return to the
measurement screen.
11.3 Damping
The DFX-8 has an adjustable damping feature, enabling the user to select from a
variety of settings according the transducer frequency and overall output. The range
of the damping feature is 50-1500 ohms. The available steps are 50, 75, 100, 300,
600 and 1500 ohms. Lower frequency transducers require less damping because
they typically have higher impedance and distort. The reverse is true for higher
frequency transducers. Nevertheless, trial and error with this feature will enable you
to find the optimal signal response and setting for your transducer. The procedure to
change the DAMPING is outlined below:
For quick access and editing, refer to navigating the Hot Menus page 22.