Dakota Ultrasonics DFX-8 plus MANUAL1 User Manual
Page 44
Dakota Ultrasonics
40
The gauge will require a few basic adjustments in order to get underway with
the calibration process. For this example, in order to avoid multiple steps for
setting the gauge features/options, the default setup will be loaded as a starting
point. However, the user should invest the necessary time getting to know the
features and options included in the DFX-8, as well as gaining a good
understanding of their function:
Loading Default Setup
1) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right and the ESC key multiple times to tab left until
the SEUTP menu is highlighted and displaying the submenu items.
Alternatively, press the
LEFT and RIGHT arrow keys multiple times
using the left button wheel.
2) Use the
UP and DOWN arrow keys multiple times on the left button
wheel to scroll through the sub menu items until DEFAULT SETUP is
highlighted.
3) Press the ENTER key to display the CONFIRM screen.
4) Press the OK key to load the DEFAULT SETUP.
Setting the Range
1) Using the left button wheel, press the
LEFT and RIGHT arrow keys to
scroll through the 9 individual Hot Menus, and the
UP and DOWN arrow
keys to scroll through the Hot Menu items, until the RANGE cell is
highlighted. Alternatively, the MEAS & ESC keys can be used to scroll
through the Hot Menu items.