Dakota Ultrasonics DFX-8 plus MANUAL1 User Manual
Page 33
![background image](/manuals/557264/33/background.png)
DFX-8 Series Ultrasonic Flaw Detectors
29
Setting Delay
1) Using the left button wheel, press the
LEFT and RIGHT arrow keys to
scroll through the 9 individual Hot Menus, and the
UP and DOWN arrow
keys to scroll through the Hot Menu items to highlight DELAY. Alternatively,
the MEAS & ESC keys can be used to scroll through the Hot Menu items.
2) Press the
LEFT and RIGHT
arrow keys multiple times on the right
button wheel to scroll the highlighted value.
3) Alternatively, press the ENTER key to display the Digits Edit Box.
4) Press the
UP and DOWN arrow keys on the left button wheel to scroll
the highlighted value.
5) Press the
LEFT and RIGHT arrow keys multiple times using the left
button wheel to scroll the digit locations.
6) Repeat steps 4 & 5 until the DELAY value is correctly displayed.
7) Press the OK key to return to the measurement screen, or ESC to cancel
entering the DELAY.
Setting Range
The Range is the distance from the delay that will be viewable on the screen. It can
be adjusted to any value greater than the Delay. The procedures to adjust the range
are outlined below: