1) Press the

UP and DOWN arrow keys multiple times on the left button

wheel until REFERENCE GAIN is highlighted, followed by pressing the

LEFT and RIGHT

arrow keys on the right button wheel to toggle on. With all

the parameters set in the previous section, press the MEAS key to return to

the main measurement screen.

Note: The user should now see a highlighted cursor displayed in the waveform

area, starting at the left most side of the display. Also note that the DGS BOX

will be highlighted in the hot menus on the right of the display.

2) Couple the transducer to the first calibration standard, and peak up on the

reflector, adjusting the gain to keep the signal height within the boundaries

of the waveform display area, and at a suitable height to allow for

attenuation at greater depths in the overall range to be tested.

3) Press the

LEFT and RIGHT

arrow keys on the right button wheel to

move the cursor over the waveform that will be used to set the

REFERENCE GAIN and draw the DGS curve.