DFX-8 Series Ultrasonic Flaw Detectors

35

until the GT1 menu is highlighted and displaying the submenu items.

Alternatively, press the

LEFT and RIGHT arrow keys multiple times

using the left button wheel.

2) Use the

UP and DOWN arrow keys multiple times on the left button

wheel to scroll through the sub menu items until ALARM is highlighted.

3) Press the

LEFT and RIGHT

arrow keys on the right button wheel to

toggle ALARM to ON. The menu items should now become active and

adjustable, as the gate is now activated.

Set Detect

1) Press the

UP and DOWN arrow keys on the left button wheel until

DETECT is highlighted.

2) Press the

LEFT and RIGHT

arrow keys on the right button wheel to

toggle the ABOVE or BELOW.

Note: Above will trigger the alarm when a signal meets or exceeds the height

of the gate, and below will detect on anything that meets or is below the height

of the gate.

Set Gate1 Start

1) Press the

UP and DOWN arrow keys multiple times on the left button

wheel until GATE1 is highlighted, followed by pressing the

LEFT and

RIGHT

arrow keys on the right button wheel to scroll the value. When the

correct start position is displayed, proceed to the next menu item setting.

2) Alternatively, press the ENTER key to display the Digits Edit Box.

3) Press the

UP and DOWN arrow keys multiple times on the left button

wheel to scroll the highlighted value.