Dakota Ultrasonics DFX-8 plus MANUAL1 User Manual
Page 80
![background image](/manuals/557264/80/background.png)
Dakota Ultrasonics
76
Note: You must set 3 points to draw a TCG curve, and a total of 16 points can
be set. Any point can be cleared by pressing the CLR key when DRAW is
active.
6) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right, and the ESC key multiple times to tab left,
until the TCG menu is highlighted and displaying the submenu items.
Alternatively, press the
LEFT and RIGHT arrow keys multiple times
using the left button wheel.
7) Press the
UP and DOWN arrow keys multiple times on the left button
wheel until DRAW is highlighted, followed by pressing the
LEFT and
RIGHT
arrow keys on the right button wheel until OFF is displayed. The
TCG curve is now set.
8) Optionally, the TCG curve can be displayed by pressing the
UP and
DOWN arrow keys multiple times on the left button wheel until CURVE is
highlighted, followed by pressing the
LEFT and RIGHT
arrow keys on
the right button wheel toggle ON.
9) Press the MEAS key to display the main measurement screen, and the TCG
curve will be displayed.