1 fuses, 2 customer supply fusing, 3 power settings – Carbolite ABA 7/35B User Manual
Page 70
70
MF60
– 1.04
26.1 Fuses
The following fuse types are present:
Supply Fuses: 32mm x 6mm type F, 16A
– 6 total (4 on board EMC filter units; 2 in separate
holders).
Auxiliary Fuses (on board EMC filter unit): 20mm x 5mm glass type F, 2A.
26.2 Customer Supply Fusing
High break capacity fuses should be used. Avoid fast-blow fuses and magnetic trip circuit
breakers - consult Carbolite if in doubt.
The supply fuse rating should be as follows:
26.3 Power Settings
The ABA 7/35 uses electronic power limiting to enable the same heating elements to be used
over the complete range of voltages from 200V to 240V.
It is important to set the voltage setting as described in section 19.
Table 26.2:1 - Mains Supply Fuses
Phase/Volts
Supply Fuse
Rating
1-phase models, 200V to 240V
40A
3-phase with neutral, 380/220V to
415/240V
16A per phase
3-phase delta, 200V to 240V
25A per phase