Fault Finding

Disconnect the equipment from the supply before carrying out any internal
investigation.

The Furnace Main Chamber or After-burner Does Not Heat
The Heat
lamp/s is/are
On

The heating elements may

have failed

Check also that the SSRs are
working correctly

The Heat
lamp/s
is/are Off

The controller shows a

zero temperature

The thermocouple may have
broken or may have a
wiring fault

The controller shows a

low temperature

The door switch may be faulty

A relay may be faulty

The SSR could be failing to
switch on due to
internal failure, faulty logic
wiring from the
controller, or faulty controller

The controller is not

illuminated

Check the supply fuses and
any fuses
in the control compartment

The controller may be faulty
or not
receiving a supply due to a
faulty
switch or a wiring fault

The Main Chamber or After-burner is Slow to Heat Up

One of the
Heat lamps
does
not light

A fuse may have
blown

Check the Supply fuses

The Heat
lamps all light
up

An element may

have failed

Check that the SSRs are
working correctly

Furnace Overheats

The heater
lamps never
turn off

The controller
shows a very high
temperature

The controller may be faulty

The controller
shows a low
temperature

The thermocouple may
have been shorted
out or may have been
moved out of the
heating chamber

The thermocouple may be
mounted the
wrong way round

The controller may be faulty

Continued