5 ABA Operation

This section describes how to heat the chamber up to temperature, and how
the automatic door locking system operates.

5.1 Switches

– Control Panel

The main control switches are on the left hand side of the control panel (see Fig 5.1.1).

5.1.1

Fig - Mains switch panel

5.1.2

Fig - Electrical shock warning lights

The

‘furnace switch’ cuts off power to the

controllers, other devices (including the balance)
and heating circuit contactors.

The ‘safety circuit test switch’ is
used to check the operation of
the two red warning lamps
referred to in section 4.1 (see
Figs 5.1.1 and 5.1.2.)

The heater lamps indicate when the power control relays are on; they flash while the

chamber is heating or maintaining temperature. They are also used to diagnose faults

– see

section 25.

5.2 Door Operation and Status Display

In previous models ‘door open’ and ‘door locked’ indicator lamps were positioned above the
‘Electric shock risk’ lamps on the upper right front of the furnace. These door status indicators
are now included (see Screen-views 4.4.1 and 5.5.1) in the status bar at the top of the
touchscreen display panel.

furnace power
switch

printer on/off
switch

safety circuit
test switch

main chamber
heater lamps

USB data port

after burner
heater lamp