N 7.3, 3 pre test requirements, 4 displayed data updates – Carbolite ABA 7/35B User Manual
Page 23
23
MF60
– 1.04
7.3 Pre Test Requirements
When the
‘Run Test’ button is pressed the controller will make several checks on the state of
the system, which must pass, before the controller will allow the test to start. The checks and
errors that are reported if a check fails are listed in Table: 7.3.1.
7.3.1
Table: Pre Test Requirement Checks
Check
Error Message
Main chamber heater power is
ON
Main Chamber heater power is switched off
After burner heater power is ON After Burner heater power is switched off
A recipe has been selected
Select recipe from library
Furnace door is closed
Test not started due to open door
Sample weight > zero
Enter Weights
Tray weight > zero
Enter Weights
Temperature set point is at the
required test temperature
Recipe temperature set point is zero
Standard recipe calibration
factor is > zero
Recipe Calibration Factor is zero
Door lock operates
Door has failed to lock
7.4 Displayed Data Updates
When the test starts the items displayed on the bottom half of Home screen will change from
temperature data (Screen-view: 7.4.1) to weight data (Screen-view 7.4.2). If a test is cancelled
then the display reverts to the temperature data display. If a test completes normally the display
remains showing the weight data until the operator presses Clear Results (ref. Table: 7.4.3).
7.4.1
Screen-view: Pre-test Display