Carbolite ABA 7/35B User Manual

Page 67

background image

67

MF60

– 1.04

Furnace Overheats (continued)
The Heat
lamp/s remain
on when the
controller is off

The SSR has

failed

‘ON’

Check for an accidental wiring
fault which
could have overloaded the
SSR

A temperature
that is higher
than the
set-point is
shown during
the test

No fault. Burning bitumen

causes the temperature to
increase above the set point.
This is normal.

Chimney Emits Smoke
Some light
smoke is
emitted
in the middle of
the test

► No fault

Dense smoke is
emitted

► The after-burner elements

or control may have failed

Investigate or contact
Carbolite

► A fuse may have blown

Check the Supply fuses

Warning Lamp & Door Faults
Red Electric
Shock Risk
lamp
lights when
door is opened

Failed cut-off relay SR1

or SR2

DO NOT USE

until relay

replaced

Fuse for left side of

chamber blown

Check and replace fuse

Red Electric
Shock Risk
lamp
does not light
when tested

Faulty lamp

DO NOT USE until lamp
replaced

Faulty

relay SR3 or SR4

DO NOT USE

until relay

replaced

Faulty momentary switch

DO NOT USE until switch
replaced

Faulty door switch

DO NOT USE until door
switch
replaced

Door does not
unlock

Furnace may have been

turned off during a test or
there may have been a
power failure during a test

Enter START PROCESS then
cancel the test.
THE RELEASE CODE ‘728’ is
REQUIRED

Failure of interface board

Loose connection lead

Check all leads