beautypg.com

N 7.7, 7 cancelling a test, 8 weight entry – Carbolite ABA 7/35B User Manual

Page 26

background image

26

MF60

– 1.04

7.7 Cancelling a Test

A test can be cancelled by pressing the

‘Cancel Test’ button, shown in the

‘In Test Display’ screen (see 7.4.2). A pass number must then be entered.
The pass number should be communicated only to authorised
operators: this is a safety feature because of the danger of opening the
door part way through a test (the operator should know the danger:

others may not). The password is fixed at 728.

It is safe to cancel a test by this method if there is no sample in the chamber (e.g. if a

‘test’ has

been started by accident), or if the sample temperature is less than 120°C. At other times it is
not safe to cancel a test, and the test should be allowed to go to completion even if the results
are not required.

7.8 Weight Entry

The ‘Weight Entry Screen’ is accessed directly from the Home screen using the ‘Enter Weights’
button. Weight entry for the tray and sample can be entered manually or automatically using an
external balance. Weights must be entered before attempting to run a test.

7.8.1

Screen-view: Weight Entry Screen