19 Reconfigure Voltage Settings For Alternative Electrical Supply

Caution: Do not increase the power limit value to a value above the design level
for the operating voltage as shown in the table below.

In the default configuration a ‘Supervisor’ level security access is required. The
power limit is a feature which is used to match the heating elements to the

supply voltage. If the voltage is stated at the time of order, no adjustment is necessary.
Adjustment is required if the equipment is relocated to a different voltage power supply.

Adjustment of the power limit should not be made arbitrarily (for example

‘to increase the

power

’) – the heating elements could burn out, or heating could become slow.

The voltage setting screen, ref. Screen-view: 19.1.1 is accessed via:

Home Screen

→ Main Menu → System Configuration → Voltage Setting

The voltage setting can be changed by pressing one of the five selection buttons. The current
setting is indicated by the green LED.

19.1.1 Image: Voltage Setting Screen