12 external balance configuration – Carbolite ABA 7/35B User Manual
Page 28
28
MF60
– 1.04
The balance settings are shown in Table 7.10.2
7.10.2
Table: Ohaus Balance Settings
Auto Print
Off
Print Internal
Off
Numeric Data
Off
Units: Gram (g)
On
Baud Rate
9600
Parity
None
Data Bits
8
Stop Bits
1
7.11 Connecting an Ohaus Explorer Pro
™ External Balance
This section refers to the Ohaus balance with a colour touch screen interface.
The external balance type must be set to Ohaus Explorer
™ in the General Settings screen, ref.
section 7.12.
A standard straight through RS232 cable is used to connect the balance to the 9 way female
RS232 connector at the rear of the furnace.
The balance settings are shown in Table 7.10.2 (same settings as for Ohaus Explorer).
7.12 External Balance Configuration
The external balance type must be configured in the General Settings screen, ref.
Screen-view:·7.12.1
This screen is accessed via:
Home Screen → Main Menu → System Configuration → General Settings
7.12.1
Screen-view: External Balance Type Configuration