7.9 Connecting an Adam Equipment

™ External Balance

Automatic weight entry requires an external balance to be connected to the 9 way female
D-type connector mounted at the rear of the furnace. To connect to an Adam Equipment

balance, use a 9 way, null modem, female to male cable. The required pin connections are 2-3,
3-2, 5-5, all other pins are unused. The external balance type must be set to Adam Equipment
in the General Settings screen, ref. section 7.12.

The balance settings are shown below:

Baud Rate

9600

Parity

None

Stop Bits

1

Number of Bits

8

Refer to the Adam Equipment balance user manual for instructions on setting these
parameters.

7.10 Connecting an Ohaus Explorer

™External Balance

This section refers to the Ohaus Explorer balance with a membrane key pad and monochrome
display.

The external balance type must be set to Ohaus Explorer

™ in the General Settings screen, ref.

section 7.12.

To connect an Ohaus Explorer

™ balance a non-standard RS232 cable with the connections

shown in Fig 7.10.1 must be used. The cable connects the balance to the 9 way female D-type
connector at the rear of the furnace.

7.10.1 Fig - Ohaus Explorer Serial Cable Connections