Pneumatic high-speed pressure controller cpc3000, Removal of the system – WIKA CPC3000 User Manual
Page 59
Pneumatic High-Speed Pressure Controller CPC3000
WIKA Operating Instruction Pneumatic High-Speed Pressure Controller ∙ Version 1.1
59
11498171.01 08/2009 GB
11. Removal of the system
Work on electrical or pneumatic/hydraulic equipment must only be carried
out by qualified and authorized service personnel, observing the corre-
sponding safety regulations.
When dismantling the system proceed as follows:
1. Make sure that there is no positive or negative pressure on the system and that all parts of
the instrument are at room temperature.
2. Switch the system off by pressing the power switch located at the rear of the instrument.
3. First pull out the power cable from the power socket and then from the power supply input.
4. Disconnect the pressure connections.
If a connection to a SWAGELOK
®
- connection is to be disconnected,
overturning or loosening the SWAGELOK
®
- connection must be prevented
with suitable tools.
5. Remove the system as necessary.
6. Protect connections with the protective caps supplied.
Warning
Warning