beautypg.com

Pneumatic high-speed pressure controller cpc3000 – WIKA CPC3000 User Manual

Page 10

background image

Pneumatic High-Speed Pressure Controller CPC3000

WIKA Operating Instruction Pneumatic High-Speed Pressure Controller ∙ Version 1.1

10

11498171.01 08/2009 GB

3. Product description

3.1 Proper use

The Pneumatic High-Speed Controller model CPC3000 is a bench top or rack mounted Digital

Pressure Calibrator/Controller used for test and calibration of mechanical pressure gauges,

pressure switches, sensors, transducers, transmitters and any pressure related devices where

time to set point is the most critical requirement.

Only dry clean air or nitrogen should be used as the pressure medium.

Shop air should be avoided and corrosive, oxidizing, condensing, explosive

gasses should be strictly avoided.

The maximum permissible supply pressure at the SUPPLY port should be

10 % over full scale value of the sensor installed.

Very fast pressure changes can damage the sensor, due to mechanical

stress on the sensor; especially if the fast pressure change leads to an

internal pressure which is higher then the full scale of the internal sensor

(even if it is only for a fraction of a second). In some cases, the internal

relief valves cannot react quickly enough to protect the sensor.

The Internal pressure sensors have a calibration certificate (see enclosure:

calibration certificate) for the entire measuring chain.

The system is not suitable for use in areas with an explosion hazard.

If the CPC3000 is not used according to this manual, safe operation of the

system is not guaranteed.

The user of the system and not the manufacturer is responsible for all

personal and material damage resulting from improper use!

Warning

Warning

Notice