Pneumatic High-Speed Pressure Controller CPC3000
WIKA Operating Instruction Pneumatic High-Speed Pressure Controller ∙ Version 1.1
43
11498171.01 08/2009 GB
8.5.1 Mensor Command Set
Command
Data
Response/Function
?
See Table Below
Returns data per the current output
format.
Acquire?
15 char string.
Ex:
Acquire? Test_stand_1
Returns:
CCC
This command is used when multiple
computers would like to control the
instrument.
Yes if acquisition is successful, No if
instrument is being controlled with
another
computer.
CCC… = name of controlling computer
See: Release? and Unlock
Address
0-31
Sets the GPIB Address.
Address?
Returns the GPIB Address.
A?
Returns the A channel pressure
reading.
AR?
Returns the A channel rate.
ARS?
Returns the A channel rate stable flag.
AS?
Returns the A channel stable flag.
Autorange
On or Off
Sets whether the auto range function is
enabled or disabled.
Autorange?
Returns whether the autorange
function is enabled or disabled.
Baro?
Returns reading from barometric
sensor.
Caldisable
Yes,no
Sets whether or not calibration of the
active sensor is disabled.
Caldisable?
Returns whether or not calibration of
the active sensor is disabled.
Cerr
None
Clears the error queue.
CID?
Returns the ID string of the regulator
for the active channel.
Cmdset
Mensor, DPI510, DPR60c, SCPI
Activates remote command set for
instrument emulation modes.
Cmdset?
Returns active command set identifier.
Control
Instrument placed in Control Mode.
Control?
Returns Yes if instrument is in control,
No if otherwise.
Crate
Slow, Medium, Fast
Sets the control rate.
Crate?
Returns the control rate – CCCC is
variable in length and corresponds
to the parameters for the CRATE
command.
Ctype?
Returns the type of regulator for the
active channel.