Pneumatic High-Speed Pressure Controller CPC3000

WIKA Operating Instruction Pneumatic High-Speed Pressure Controller ∙ Version 1.1

45

11498171.01 08/2009 GB

Command

Data

Response/Function

IP?

nnn.nnn.nnn.nnn

Returns the IP address of the instru-
ment.

Keylock

Yes or No

Locks or unlocks keyboard.

Keylock?

(Yes or No)

Returns Yes or No.

List?

Pri,X,X;Sec,X,X;Bar,1

Returns list of available turn-downs on
installed sensors in the active channel.
X will be non-existent if the turndown
isn’t available.

Listrange?

PRI,1, min,max,2, min, max;SEC,1,
min,max,2, min,max,Bar,min,max

Returns the ranges of the installed
sensors for the active channel.

Localgravity

Value in ft/s²

Sets the local gravity in feet/sec².

Localgravity?

n.nnnnne+nn

Returns the local gravity in feet/sec².

LowerLimit

Value inside primary xducer Range
on turndown #1 in current units.

Sets the lower control limit for the
instrument.

LowerLimit?

xxxxxxx

Returns the lower control limit for the
instrument in current units.

Measure

None

Instrument placed in Measure Mode.

Measure?

(Yes or No)

Returns YES if instrument is in
measure, No if otherwise.

Netmask

nnn.nnn.nnn.nnn

Sets the Ethernet network mask.

Netmask?

nnn.nnn.nnn.nnn

Returns the Ethernet network mask.

Outform

1 to 8 – see table below

Sets the output format.

Outform?

X

Returns the output format – see table
below

Peakmax?

n.nnnnne+nn

Returns the maximum pressure since
peakreset was sent.

Peakmin?

n.nnnnne+nn

Returns the minimum pressure since
peakreset was sent.

Peakreset

None

Resets the peak values.

Port

nnnnnn

Sets the Ethernet port of the instru-
ment

Port?

nnnnn

Returns the Ethernet port of the instru-
ment

Ptype

Absolute or Gauge

Sets the instrument pressure type
– gauge only works if the optional
barometric sensor is installed

Ptype?

CCCCC

Returns “Absolute” or “Gauge” for the
pressure type

RangeMax?

XXXXXXX

Returns the maximum range of the
active transducer and turndown in the
current units.

RangeMin?

XXXXXXX

Returns the minimum range of the
active transducer and turndown in the
current units.

Rate?

XXXXXXX

Returns the rate reading of the instru-
ment in current units/second.