beautypg.com

Pneumatic high-speed pressure controller cpc3000, Specifications, Specifications cpc3000 – WIKA CPC3000 User Manual

Page 17

background image

Pneumatic High-Speed Pressure Controller CPC3000

WIKA Operating Instruction Pneumatic High-Speed Pressure Controller ∙ Version 1.1

17

11498171.01 08/2009 GB

4. Specifications

Specifications

CPC3000

Pressure ranges

bar

-1 ... +70 (depending on sensor)

Pressure types

Absolute, gauge or bi-directional ranges

Precision

% FS

< 0.015

Accuracy

% FS

< 0.025

Pressure units

selectable and freely definable

Control stability

% FS

< 0.004

Slew rate

sec.

< 3 (with a sudden pressure increase of 10 % FS in a 150 ml test volume)

Control range

% FS

0 up to 100

Test volume

ccm

50 ... 1,000 (without throttle)

Pressure ports

mm

4 ports with 7/16"- 20 F SAE. incl. 6 mm tube fitting adaptors

Filter elements

40 micron filter element included in each pressure port

Permissible pressure media

clean dry non-corrosive gases

Wetted parts

aluminium, brass, 316 and 316L stainless steel, Buna N, FPM/FKM,
glass filled epoxy, RTV, Nylon, Ceramic

Overpressure protection

safety relief valve

Instrument version

desk top/optional: incl. rack mounting kit

Screen

7.0" colour LCD with touch-screen

Resolution

digits

4 ... 6

Internal measuring rate

1/sec

25

Display update

1/sec

4

Warm-up time

min

approx. 15

Digital Interface

Ethernet, IEEE-488, USB

Command sentences

Mensor, SCPI, others optional

Power Supply

V AC

100 … 240, 50-60 Hz

Energy Consumption

VA

maximum 90

Permissible pressure

Supply port

% FS

~ 110

Measure/Control port

% FS

max. 105

Permissible

Operating temperature

°C

10 … 50

Storage temperature

°C

0 ... 70

Air humidity

%

0 … 95 (relative humidity without moisture condensation)

Operating position

Horizontal or slightly tilted

Compensated temperature
range

°C

15 …. 45

Weight

kg

approx. 9.1

Dimensions

mm

see technical drawings

CE-mark

Conformity certificate

Calibration*

Incl. factory calibration certificate 3.1 per DIN EN 10 204

* Calibration in a horizontal position.