5.5 Rear panel

Four pneumatic pressure ports are located on the rear panel (see below: "Figure - Rear
panel"). Positioned on the left are the ethernet and RS-232 and GPIB connector, the off/on
switch, the line fuses, and a protective grill covering the ventilating fan.

5.6 Pressure connections

The pressure connections must be installed according to the following

instructions, observing the relevant regulations. The installation should be

performed by persons familiar with, and who can work according to, the

safety regulations for working on pneumatic/hydraulic systems.

When making up a connection to an o-ring adapter port use a back-up wrench
to prevent over-stressing the threads in the manifold block.

All of the pressure ports on the rear are female 7/16 - 20 SAE/MS straight threads per
MS16142 and SAE J514 table 14.
They require a tube fitting boss seal with an o-ring per MS33656.
WIKA provides female 1/4 inch and 6 mm tube fittings with the instrument.
The pressure connections can be made to these adapters with the proper mating hardware.
Do not use sealant on fittings sealed with an o-ring.

Figure - Rear panel

Fan

Mains input socket

Micro fuse

Power supply

Supply Port

Measure/Control Port

Reference Port

Vent

Exhaust/Vacuum Port

IEEE-488.2

Warning

Notice

310

148

191

228

259

Ethernet USB