Pneumatic high-speed pressure controller cpc3000 – WIKA CPC3000 User Manual
Page 46
Pneumatic High-Speed Pressure Controller CPC3000
WIKA Operating Instruction Pneumatic High-Speed Pressure Controller ∙ Version 1.1
46
11498171.01 08/2009 GB
Command
Data
Response/Function
Rdecpt?
Returns the number of rate decimal
points for the active channel. (see
Resolution)
Release?
15 char string.
Ex:
Release? Test_stand_1
Returns:
CCCcr>
This command is used to release
control of the instrument in a multiple
computer environment.
Yes if release is successful.
No if instrument is being controlled
with another computer.
CCC… = name of controlling computer
or AVAILABLE
See: Acquire? and Unlock
Resolution
n
Sets the number of significant digits.
See decpt.
Resolution?
Returns the number of significant
digits. See decpt.
Rfilter
Value in %
Sets the % of the rate filter
Rfilter?
Returns the rate filter.
Rsetpt
Value in current units
Sets the rate setpoint
Rsetpt?
Returns the rate setpoint
Rfreq
Value in frequency
Sets rate Butterworth corner frequency.
Rfreq?
Returns rate Butterworth corner
frequency.
Rwindow
Value in current units
Sets rate exponential filter window.
Rwindow?
Returns rate exponential filter window.
Sbaud
9600, 19200, 38400, 57600
Sets the serial baud rate.
Sbaud?
Returns the serial baud data.
Sdata
7 or 8
Sets the serial data bits.
Sdata?
Returns the serial data bits number.
Sensor
C, X
Sets the active sensor where C =
Primary or Secondary and X is the
turndown.
Sensor?
Returns active sensor as above.
Sensorid?
V.VV
Returns the active sensor’s serial
number and firmware version.
Setpt
value inside upper and lower limits
and inside the range of the active
sensor and turndown.
Sets the control setpoint for the instru-
ment.
Setpt?
Returns the control setpoint in current
units.
Setpt%
Value in % of current range
Sets the control setpoint in % of
current range.
Setptpct
Value in % of current range
Sets the control setpoint in % of
current range.
Setptpct?
Returns the current setpoint in % of
current range.