Pneumatic high-speed pressure controller cpc3000 – WIKA CPC3000 User Manual
Page 47

Pneumatic High-Speed Pressure Controller CPC3000
WIKA Operating Instruction Pneumatic High-Speed Pressure Controller ∙ Version 1.1
47
11498171.01 08/2009 GB
Command
Data
Response/Function
Span
desired pressure or ?
Sets span on active transducer or for
?, clears previous value, must be
> 50 %FS and has a 1 % limit.
Span?
Returns span scale factor for active
transducer.
Sparity
Even, Odd, None
Sets the serial parity.
Sparity?
Returns the serial parity.
Sstop
1 or 2
Sets the serial stop bits.
Sstop?
Returns the serial stop bits.
Stable?
Returns YES if instrument is stable, or
No.
Stabledelay
0 to 65535
Sets the stable time to the number of
seconds specified.
Stabledelay?
Returns the stable time.
stabletime
0 to 65535
Sets the stable time to the number of
seconds specified.
Stabletime?
Returns the stable time.
StableWin
%fs value
Sets the stable window as a %FS for
the active sensor and turndown.
StableWin?
Returns the stable window.
Standby
None
Instrument placed in Standby Mode.
Standby?
Returns Yes if instrument is in Standby,
No if otherwise.
Step
value inside upper and lower limits
and inside the range of the active
sensor and turndown.
Sets the control step size for the instru-
ment.
Step-
Optional value
Jogs the setpoint down one step.
Step+
Optional value
Jogs the setpoint up one step.
Step?
Returns the control step for the instru-
ment.
Step%
Value in % of current range
Sets the control step in % of current
range.
Steppct
Value in % of current range
Sets the control step in % of current
range.
Steppct?
Returns the current step in % of
current range.
Units
units code or text in table below
Sets the instrument engineering units.
Units?
Returns the instrument units in a text
string.
Unlock
None
Releases Acquire locks. See Acquire?
and Release?
UpperLimit
Value inside primary xducer Range
on turndown #1 in current units.
Sets the Upper control limit for the
active transducer.
UpperLimit?
Returns the upper control limit for the
active transducer.