Pneumatic high-speed pressure controller cpc3000 – WIKA CPC3000 User Manual
Page 50
Pneumatic High-Speed Pressure Controller CPC3000
WIKA Operating Instruction Pneumatic High-Speed Pressure Controller ∙ Version 1.1
50
11498171.01 08/2009 GB
Command
Data
Response/Function
_pcs4 stabledelay?
Returns the number of readings
that must be within the stable
window before a stable pressure is
indicated.
_pcs4 stablewindow
Sets the pressure window that is
used to indicate pressure is stable.
_pcs4 stablewindow?
Returns the pressure tolerance
allowed for a stable pressure
indication as a % of span of the
active transducer.
_pcs4 stat?
Returns Mode and stable flag
status “mode, stable CR LF”.
_pcs4 unit
Sets the instrument to specified
engineering units.
_pcs4 unit?
Returns the current engineering
units and the type of transducer (A,
G, D).
_pcs4 xducer?
Returns the number of the currently
active transducer.
_pcs4 xducerid?
_pcs4 zero?
Returns the stored zero offset of
the active transducer & turndown in
the current pressure units.
8.5.3 PCS 200 Commands Emulated
Command
Data
Response/Function
CX
Control Pressure at last control point
and units
C$nnnnnnnX
Control Pressure at n in units $
C$nnnnnnnsX
Control Pressure at n in units $
D#X
CAL POINT CONTROL MODE NOT
SUPPORTED
EX
Clear Error/Clear Service Request
E?X
Return error code and clear error
F$nnnnnnn1X
Re-initialize; $, n ignored
F$nnnnnnn2X
RETURN CAL DATA NOT SUPPORT-
ED
F$nnnnnnn3X
Return Unit ID string; $,n ignored
F$nnnnnnn5X
RETURN QPS TEMPERATURE NOT
SUPPORTED
F$nnnnnnn6X
RETURN NULL METER READING
NOT SUPPORTED