beautypg.com

Pneumatic high-speed pressure controller cpc3000 – WIKA CPC3000 User Manual

Page 26

background image

Pneumatic High-Speed Pressure Controller CPC3000

WIKA Operating Instruction Pneumatic High-Speed Pressure Controller ∙ Version 1.1

26

11498171.01 08/2009 GB

7.3 Setpoint Entry

The control setpoint can be entered using the default numeric keypad or the alternate STEP
or JOG keypads that appear on the right side of the main menu when selected using the
[SETPOINT ENTRY SIDE-MENU] tabs

on the lower right hand side of the

main menu. These alternative methods of entering the setpoint have advantages in different
situations and have been designed to increase ease of use and productivity.

The numeric keypad shown in "Figure - Numeric Keypad" to the left,
is the default keypad that appears every time the unit is turned on or
can be activated using the

[0-9] tab. A setpoint value can be entered

directly using this keypad. As the value of the setpoint is entered the
setpoint field will turn blue and the entered value will appear in the field.
The Setpoint value can be deleted completely using the

[DELETE]

key, the last digit of the entered setpoint can be deleted using

the

[CLEAR ENTRY]

key or the setpoint value can be accepted

using the

[ACCEPT ENTRY]

key. When the [ACCEPT ENTRY] key

is pressed the setpoint field will turn white and the new setpoint will
become active. In control mode, the controller output will ramp to the
entered setpoint.

Caution: If the [ACCEPT ENTRY] key is not pressed

the previously entered setpoint will remain active.

The STEP keypads shown in "Figure - STEP
Keypads", provide a way to increment the
setpoint by defined steps. Steps are a percent of
the user defined limits set in the

MAIN -> SETUP

-> CONTROL or actual pressure values that are
displayed in the pressure units selected in the
main menu. A total of 12 steps are provided.
When a step is pressed the related setpoint is
immediately entered as the active setpoint. In
control mode, the controller output will ramp to
this setpoint. The STEP Keypad can be modified
in the

MAIN -> SETUP -> DISPLAY menu

discussed in chapter 7.8.1.

The JOG Keypad shown in "Figure - JOG Keypad", provides a way to
jog the setpoint up or down by small increments. The increments are
determined by the resolution, the maximum control limit and/or the
units of measure. For example, if the resolution is set to display four
decimals then the small triangle pointing up will change the setpoint by
+0.0001 and the small triangle pointing down will change the setpoint
by -0.0001. In the same way, the medium triangles will change the
setpoint by ±0.0010 and the large triangles will change the setpoint by
±0.0100 as shown in the figure. When the Resolution, the maximum
control limit or the units of measure are changed so that three decimals
are displayed, then the JOG functions will change to ±0.001, ±0.010
and ±0.100 respectively. This is useful when adjusting the pointer of a
manometer congruent to the skale.

Figure - Numeric Keypad

Figure - STEP Keypads

Figure - JOG Keypad