Swapping devices – Vaisala viewLinc-4.1-M211518 User Manual
Page 53

Chapter 3: System Settings
42
©
2012 Vaisala Canada Inc
3
Enter the following:
a
Disconnect After Scan: To maintain a persistent con-
nection to the device, choose False.
b
Connection Type: If the transmitter has a LAN or
WLAN module, select Network. If you use an external
Digi device, choose COM Port.
c
IP Address: See your administrator or check the De-
vice box.
d
Sample Rate: Choose 90 (recommended, but you can
modify the rate if you need more or less recorded da-
ta).
e
Connection Timeout: To ensure continuous monitor-
ing, do not change. Contact your network administra-
tor if required.
4
To add several 300 Series Transmitters, HMT140 Series
Transmitters or a combination of Vaisala Veriteq devices
and 300 Series Transmitter devices, select Upload
Definitions File then enter or browse to the correct file. To
create a Definitions File, see the FAQ on page 145.
Swapping Devices
If you need to send a device for maintenance (such as
recalibration), you will need to swap it for another device to
ensure continuous monitoring and alarming.
Note:
If a device is removed without being swapped,
this will trigger a communication alarm.
When a device is swapped, the change is noted on the
Location History report (the report shows the device serial
number for a reporting period). If, during the reporting period,
the device was swapped, this event is listed in the report
summary.
To swap a device:
1
Ensure the new device has the same settings as the
device to be swapped (see
Options | System
Configuration | Locations Manager
, or refer to your
Vaisala product User Guide).